User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 7562A:Log Voltmeter/Converter DATE: 10-Feb-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 37 CONFIGURATION: HP 3325A STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 IEEE [@3325] AM .0005 VR 1.002 ASK- R P F W 1.003 ASK+ X 1.004 STD Stanford Research FS700 1.005 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.006 DISP Allow the UUT a 20 minute warm-up. 1.007 DISP ÿ Connect the 5700A and the UUT as follows: 1.007 DISP ÿ [27][91]1m 5700A TO 7562A 1.007 DISP ÿ OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ> INPUT (HI) 1.007 DISP ÿ OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ> INPUT (COM) 1.008 DISP ÿ Connect the UUT and the 1281 as follows: 1.008 DISP ÿ [27][91]1m 7562A TO 1281 1.008 DISP ÿ OUTPUT HI ÄÄÄÄÄÄÄÄÄ> INPUT HI 1.008 DISP ÿ OUTPUT (COM) ÄÄÄÄÄÄÄÄÄ> INPUT LO 1.009 DISP Set the UUT front Panel controls as follows: 1.009 DISP function switch ..................................... DC 1.009 DISP RANGE ........................................ 1mV - 10V 1.010 CALL HP 7562A Sub 1 1.011 CALL HP 7562A Sub 2 1.012 HEAD {} 1.013 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 DISP {º THIS COMPLETES THE VERIFICATION º} 1.013 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.014 END